Installing and Operating Your Multicell Transport
Removing the Multicell Transport from the Spectrophotometer
Removing the Multicell Transport from the
Spectrophotometer
Use this procedure in case you want to remove the multicell transport from
your spectrophotometer, for example, for repairs or simply if you want to use
other cuvette holders instead.
1 Turn off the water circulation, if connected.
2 Disconnect the tubing from the water supply.
3 Disconnect the multicell transport (cell holder can now be moved from left to
right).
4 Loosen the mounting screws.
5 Lift the multicell transport out of the sample compartment.
25