Agilent Technologies 8453 Operator's Manual page 17

Uv-visible spectrophotometer
Table of Contents

Advertisement

Figure 4
Tubing clamp (C)
3 Attach tubing clamps to fittings (A and B).
4 Before circulating water through the multicell transport, test the tubing
5 Do this several times to establish that the tubing neither pulls loose nor bends
6 A maximum water pressure of approximately 3.5 bar (50 psi) can be supplied
Installing and Operating Your Multicell Transport
Installing Your Multicell Transport
Multicell Transport with Fitting Connections
arrangement by moving the cell holder to the HOME-position and cell
position 8, using your operating software.
sharply. Start the water flow and repeat this test. Make sure that there is no
restriction of the water flow.
to the thermostatted cell holder.
Inlet fitting (A)
Outlet fitting (B)