Data rates with Agilent data systems

Advanced Operation Manual
Baseline upset due to valve switch
Pause signal here
The GC can process data at various data rates, each
corresponding to a minimum peak width. The table shows
the effect of data rate selection.
Table 58
EZChrom/ChemStation data processing
Data rate, Hz
Minimum peak
width, minutes
500
0.0001
200
0.001
100
0.002
50
0.004
20
0.01
GC Output Signals
Resume signal here
Relative
Detector
noise
5
FID
3.1
FID
2.2
FID/FPD/NPD
only
1.6
1
7
Column type
Narrow-bore,
0.05 mm
Narrow-bore,
0.05 mm
Capillary
233