Programming run time events

The run table

Adding events to the run table

Advanced Operation Manual
• Detector gas flow (on/off), including NPD H
• Inlet septum purge flow
Press [Run Table].
1
Press [Mode/Type] to see the available run time events.
2
Scroll to the event to be programmed. Press [Enter].
3
Enter values for the Time: and the other parameter. Press
4
[Enter] after each entry.
Press [Mode/Type] to add another event. Press [Status] to
5
terminate entries.
The programmed events are arranged in order of execution
time in the Run Table. This is a brief example:
RUN TABLE (1 of 3)
Time:
0.10
Valve #2
On
RUN TABLE (2 of 3)
Time:
3
Analog signal 2 range 2
RUN TABLE (3 of 3)
Time:
4.20
Valve #2
Off
To add new events to the run table, press [Mode/Type]
1
while on the Time: line of any entry.
2
Select the event type.
Set appropriate Time: and other parameters. Some require
3
numbers; others require [On/Yes] or [Off/No].
Repeat until all entries are added. Events are
4
automatically placed in order by execution time.
Programming
2
Event 1 rotates a valve.
Event 2 adjusts the signal
range.
Event 3 resets Valve #2 to its
original position in preparation
for another run. Valves do not
reset automatically.
1
fuel gas
13