Lighting The Fpd Flame - Agilent Technologies 7890 Series Advanced Operation Manual

Gas chromatograph
Hide thumbs Also See for 7890 Series:
Table of Contents

Advertisement

Lighting the FPD flame

Advanced Operation Manual
Table 54
Recommended flows (continued)
Detector gases
Hydrogen
Air
Carrier + makeup
Helium, either as carrier or makeup gas, may cool the
detector gases below the ignition temperature. We
recommend using nitrogen rather than helium.
Before trying to light the flame, have the detector at
operating temperature. Removing the condensate tubing may
help, but be sure to replace it before making runs.
It is difficult to light the flame with the flows shown in
Table
53, particularly in the sulfur mode. If the flame will
not light with the sulfur mode flows shown, change to the
phosphorus mode flows. After ignition, gradually reduce the
flows to the sulfur values. Some experimentation will be
needed.
When either of the flame ignition methods in this section is
used, the FPD automatically performs this sequence:
1
Turns all detector gases—air, hydrogen, makeup—off.
Carrier remains on.
Sets air flow to 200 mL/min.
2
Turns the glow plug ignitor on.
3
Ramps the hydrogen flow from 10 to 70 mL/min.
4
Resets the air flow to the air flow setpoint.
5
Resets the hydrogen flow to the hydrogen flow setpoint.
6
Turns the makeup gas on.
7
Compares the signal change with the Lit offset value. If the
8
change is greater than Lit offset, declares the flame on
(lit). If it is less, declares the flame off (not lit).
For this process to work, there must be enough air pressure
to the pneumatics module to provide 200 mL/min flow. We
recommend a supply pressure of 90 psi.
Sulfur mode flows,
Phosphorus mode
mL/min
flows, mL/min
50
75
60
100
60
60
5
Detectors
209

Advertisement

Table of Contents
loading

Table of Contents