Selecting an NPD bead type

Advanced Operation Manual
column in the constant pressure mode, choose Column
+makeup=constant.
c
If your column is not defined, enter a makeup gas
flow. Only constant flow is available.
Monitor the offset adjustment process.
13
If Auto Adjust is On, the adjust offset process starts
a
automatically when the detector reaches setpoint. If
Auto Adjust is Off, the Bead Voltage will gradually go to
the last setpoint after the bead reaches setpoint
temperature and the Dry Bead time has elapsed.
If you need to set a new target offset, enter an Adjust
b
offset value. Adjust offset starts when the detector
reaches setpoint.
If Auto Adjust is Off, you can manually start the Adjust
c
offset process by scrolling to Adjust offset, then pressing
[On/Yes].
d
If your standard operating procedures require that you
set the bead voltage directly, see
voltage manually (optional)"
Three beads are available:
Table 50
NPD beads
Bead type
Part number
White ceramic
G1534-60570
Black ceramic
5183-2007
Blos bead
G3434-60806
Ceramic beads
Detectors
"Setting NPD bead
on page 199.
Advantages
Disadvantages
Standard
Phosphorus tails
Durable, no
Lower nitrogen
phosphorus
sensitivity,
tailing
about 40%
Moisture
resistant, long life
Blos bead
5
193