Table 6
Approximate minimum viable inlet pressures for split/splitless inlet in split mode, in psi (kPa)
Helium and hydrogen carrier gases
Split liners - 5183-4647, 19251-60540
Splitless liners - 5062-3587, 5181-8818
Nitrogen carrier gas
Split liners - 19251-60540, 5183-4647
Splitless liners - 5062-3587, 5181-8818

Selecting the correct S/SL inlet liner

Table 7
Split mode liners
Liner
Advanced Operation Manual
Split vent flow (mL/min)
50–100
2.5 (17.2)
4.0 (27.6)
3.0 (20.7)
4.0 (27.6)
These numbers are based on the resistance to flow of new,
clean inlet systems. Sample condensation in the split vent
tube or a dirty filter can make these values non- attainable.
Split liner
A good liner for split mode operation will offer very little
restriction to the split flow path between the bottom of the
liner and the inlet gold seal and between the outside of the
liner and the inside of the injection port body. The preferred
Agilent split liner, part number 5183- 4647, incorporates a
glass positioning bead on the bottom to facilitate this. It will
also incorporate glass wool or some other source of surface
area inside the liner that provides for complete sample
vaporization across the boiling point range of the sample.
Select an appropriate liner from
Description
Volume
Low Pressure Drop
870 µL
– Positioning Bead
4mm ID, Glass Wool 990 µL
Empty Pin & Cup
800 µL
Packed Pin & Cup
800 µL
100–200
200–400
3.5 (24.1
4.5 (31)
5.5 (37.9)
8.0 (55.2)
4.0(27.6)
6.0 (41.4)
Table
7.
Mode
Deactivated
Split – Fast
Yes
Injection
Split – Fast
No
Injection
Split – Manual
No
Only
Split – Manual
No
Only
3
Inlets
400–600
6.0 (41.4)
11.0 (75.4)
Part Number
5183-4647
19251-60540
18740-80190
18740-60840
49