Programming column pressure or flow

158
remains at that value for Final time 1. You can then add a
second and third ramp, each consisting of a Rate, a Final
value (pressure or flow), and a Final time.
The program ends when it reaches a Rate that is set to 0
(Off).
When a flow or pressure program is running, the Pressure,
Flow, and Velocity lines that you used to set constant
conditions show the progress of the program.
The oven program determines the length of the run. If a
flow or pressure program ends before the analytical run
does, the flow (or pressure) remains at the last final value.
Press [Col 1] or [Col 2], or press [Aux col #] and enter the
1
column number.
Scroll to Initial pressure (or Initial flow). Type the desired
2
value and press [Enter].
Similarly, enter a value for Initial time. This completes the
3
initial part of the program.
To begin a ramp, enter a positive value for Rate 1. It does
4
not matter whether you are programming up or
down—the rate is always positive.
If Rate 1 is zero, the program ends here. If you enter any
5
other value, the Final value lines for the first ramp
appear and the cursor moves to the line.
Enter values for Final pressure 1 (or Final flow 1) and Final
6
time 1. This completes the first ramp.
To enter a second ramp, scroll to the appropriate Rate
7
line and repeat steps
can be entered.
5
and 6. A maximum of 20 ramps
Advanced Operation Manual