To Configure The Npd - Agilent Technologies 7890 Series Advanced Operation Manual

Gas chromatograph
Hide thumbs Also See for 7890 Series:
Table of Contents

Advertisement

Table 52
Jets for adaptable fittings
Figure 4 ID
Jet type
1
Capillary with extended jet
(recommended)
2
Capillary
3
Capillary, high-temperature
4
Packed
70.5 mm
1
61.5 mm
2
61.5 mm
3
63.6 mm
4
Figure 2
Adaptable NPD jets

To configure the NPD

Advanced Operation Manual
51.5 mm
1
43 mm
2
43 mm
3
Figure 1
Capillary optimized NPD jets
For the adjustable NPD, select one of the following from
Table
52.
Part number
G1534-80590
19244-80560
19244-80620
18710-20119
1
In addition to the Ignore Ready and Makeup gas type, the NPD
requires the following configuration settings. Scroll to each
and enable/disable using [On/Yes] or [Off/No].
1
2
Jet tip id
Length
0.29 mm (0.11 inch)
70.5 mm
0.29 mm (0.011 inch)
61.5 mm
0.47 mm (0.018 inch)
61.5 mm
0.46 mm (0.018 inch)
63.6 mm
2
3
4
5