8
Auxiliary Devices

About Auxiliary Detectors

The GC supports up to two auxiliary detectors in addition to
the Front and Back detectors that mount on the top of the
oven.
Aux Det # 1
This can only be a TCD, and mounts in a
carrier on the left side of the oven together with its flow
module.
Aux Det # 2 This can only be an analog input board (AIB). It
is used to receive and process data from a non- Agilent
detector or other source.