Safety Practices and Hazards
Changes or modifications not expressly approved by Agilent
Technologies could void the user's authority to operate the
equipment.

ICES/NMB-001

This ISM device complies with Canadian ICES- 001.
Cet appareil ISM est conforme à la norme NMB-001 du Canada.

Flame Operation

Flammable Solvents

Unskilled, improper, or careless use of flammable solvents in or near
an atomic absorption spectrometer can create explosion hazards and
fire hazards. These can result in death, or severe personal injury or
burns.
Remember at all times that the combination of a flame and flammable
solvents can present a serious hazard. All relevant safety practices
governing the use of flammable solvents must be strictly followed.
To reduce the possibility of fire or explosion:
20
When initially selecting an organic solvent, choose a solvent
having the highest flash point consistent with your analytical
requirements.
Never use a solvent having a specific gravity lower than 0.75.
Never leave uncovered containers of flammable solvent standing
near the burner. When aspirating such solvents, always use a
covered container and feed the capillary tubing through a 2 mm
diameter hole in the cover. Always use the smallest volume of
solvent consistent with your analytical requirements.
Agilent 55B AA Spectrometer User's Guide