Gc / External Events (Unspecified, Non-Agilent Instrument) - Agilent Technologies 8860 Installation And First Startup

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

B
Cabling Diagrams and Remote Start/Stop
GC / External Events (unspecified, non-Agilent
instrument)
Number
1
Connector
24 volt control output
1
2
3
4
Relay contact closures (normally open)
5
6
7
8
66
GC/ALS
Part number and description
G1530-60590, External events cable, 8-pin/spade lugs
Signal name
Maximum rating
24 volt output 1
150 mA output
24 volt output 2
150 mA output
Ground
Ground
Contact closure 1
48 V AC/DC, 250 mA
Contact closure 1
Contact closure 2
48 V AC/DC, 250 mA
Contact closure 2
1. External event cable,
8 pin/spade lugs,
G1530-60590
Wire color
Yellow
Black
Red
White
Orange
Green
Brown or violet
Blue
Installation and First Startup
Corresponds to valve #
5
6
7
7
8
8