6.1 Stand By
From the "STAND BY" state:
▪ Press the ON/OFF button to bring the balance to working condition.
▪ Press the ON/OFF button again to return to the "STAND BY" state.
6.2 Simple weighing
Place the sample to be weighed on the plate and read the value of the weight on the display as
soon as the asterisk stability symbol ж appears.
01-06-10
13:15:24
0.0000
g
-0-
7
Calibration
The Accuris Analytical Balance carries out mass measurements using gravity (g). Differences in
geographical regions and altitudes will affect the gravitational acceleration (g).
The balance must therefore be adapted to specific environmental conditions by calibrating to
obtain accurate measurements.
7.1 Balances with external calibration
7.1.1 External calibration
The calibration is carried out through the CAL button.
1. Press the CAL button with the plate unloaded; the word CAL will be displayed.
01-06-10
13:15:24
Cal
-0-
9