Calibration Data - Accuris W3101A-120 Operation Manuals

Analytical balance series dx (graphic dot matrix display)
Table of Contents

Advertisement

5. Once the internal calibration has been stored, the balance will return to the normal weighing
condition.
6. Now re-enter the calibration menu as described in paragraph 9.1.12 and set the desired
internal, automatic, or external calibration modality.
WARNING: this procedure must be carried out only using class E2 or equivalent
reference masses.

9.13 Calibration data

This function allows you to display the data relative to the last calibration carried out.
Date
Calibration mode
Correction
1. Select the calibration data parameter as described in paragraph 10.
The data relative to the last calibration carried out will be shown on the display:
2. Press the PRINT button to print the calibration data.
3. Press the ON/OFF button to exit from the screen and return to the balance parameters menu.
It will now be possible to select another parameter or return to weighing mode by pressing the
ON/OFF button.
Calibration data
10-06-10
External calib.
Corr.:
29
2000.0g
1.2g