Continuous Transmission Mode; Ondemand Transmission Mode - Accuris W3101A-120 Operation Manuals

Analytical balance series dx (graphic dot matrix display)
Table of Contents

Advertisement

11.4.1 Continuous Transmission mode

String transmitted is composed by the following 14 characters:
▪ First character: weight sign (blank or -)
▪ Second to ninth character: weight or other data
▪ Tenth to twelfth character: weight unit symbol
▪ thirteenth character: stability indicator
▪ fourteenth character: carriage return
▪ fifteenth character: line feed
Eventual non-significative zero are put as spaces.
In the following table the various transmission formats are shown:
Weight mode (valid for both continuous and on demand transmission)
Sign
11.4.2 On demand transmission mode
When in on demand mode, the transmitted data to computer do not include only informations of
the weight value but also date/time and other informations that depend on the function you're
currently using are transmitted to computer
Below are shown the data transmitted in each situation possible:
WEIGHT:
--------------------
03-04-11 10:13:44
--------------------
Weight:
0.00 g
PIECE COUNTING:
--------------------
03-04-11 10:49:28
--------------------
Pcs.:
10
Weight:
100.02 g
MPW:
10.00 g
DENSITY:
--------------------
03-04-11 10:51:15
--------------------
d: 1.4504 g/cm3
weight
10°
11°
measure unit
55
12°
13°
14°
Stability
CR
15°
LF

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

W3100a-220Dx w3101a-120Dx w3100a-220Dx w3101a-220

Table of Contents