12 Error codes

ERROR MESSAGE
ON DISPLAY
ERR01
ERR02
ERR03
ERR04
ERR05
ERR06
ERR07
ERR08
ERR09
ERR10
ERR F
"UNLOAD"
"CAL But":
MEANING
Weight not stable after operation
of tare
impossible to start the calibration
due to instability of the balance
calibration weight not correct or
balance unstable
weight of samples for the piece
counting function not adequate or
unstable
impossible to print because of
unstable weight
Weight is not stable in density
mode
Weight is not stable in percent
weighing mode
Anomaly with autocalibration
motor
Weight is not stable in formulation
mode
Weight of component out of
tollerance in formulation mode
Flash memory damaged
weight loaded on the pan or pan
not positioned properly
the balance requires to be re-
calibrated
Over-load condition
Under-load condition
POSSIBLE SOLUTIONS
Protect the balance from air
flows or from vibrations
from the table.
Protect the balance from air
flows or from vibrations
from the table.
Calibrate with correct
weight or protect the
balance from environmental
disturbances. In models
with internal calibration
remove the screw in the left
lower part of the balance
(see part 5)
Select a larger number of
samples or protect the
balance from vibrations.
Protect the balance from
environmental disturbance
Protect the balance from
environmental disturbance
Protect the balance from
environmental disturbance
Contact service staff
Protect the balance from
environmental disturbance
Reduce quantity
Ask assistance from
authorized service staff
Remove the weight from
the pan or position properly
the pan and underpan.
Unload weights, if any, on
the pan, and press the CAL
button
Unload the weights loaded
on the pan
Place properly pan and
underpan
61