Storage Conditions - Accuris W3101A-120 Operation Manuals

Analytical balance series dx (graphic dot matrix display)
Table of Contents

Advertisement

▪ Connect the power supply to the power inlet at the back of the balance (see Section 5).
▪ Connect the power supply to a nearby outlet, which must be easily accessible. Press the
power button on the right side of the control panel to power on the balance.
▪ Balance Warm-up:
▪ It is necessary to allow the balance to equilibrate to ambient conditions before use.
Wait at least 8 hours after initially plugging in the analytical balance (0.0001g)
Then calibrate the balance as described in Section 7.
It is strongly suggested not to disconnect power cord from the balance to turn it
off. Use the ON/OFF button to put the balance into stand-by mode when not in
use.
▪ Calibrate the balance every time that it is moved to another location.
▪ Periodically check the balance's calibration.
▪ Do not drop objects of an excessive weight on the balance weigh pan, as doing so may
damage to the weigh cell mechanism.
▪ Service should be carried out by trained personnel, and spare parts used must be original.
Contact your distributor or Accuris Instruments for service.
2

Storage conditions

▪ Storage temperature +5 °C to +40°C
▪ Storage humidity 45% - 75%.
▪ Keep the original balance packaging in case the instrument needs to be shipped for
servicing; disconnect all cables and any accessories to prevent unnecessary damage.
▪ Do not expose the balance to extreme temperatures and humidity unnecessarily, and avoid
violent shocks and vibration.