user by pressing the CAL button.
Before carrying out the internal calibration, ensure that no weight is loaded on the plate.

9.12.3 External calibration (E-CAL) [locked for verified balances]

The balance will be calibrated by using the external reference mass.
(Follow the procedures described in paragraph 8.1.2)

9.12.4 Technical calibration (TEC-CAL) [locked for verified balances]

This function allows the internal reference mass to be calibrated whenever assistance-control-
maintenance interventions make this necessary.
1. After having selected the TEC-CAL calibration mode, press the CAL button with the plate
empty. The word "CAL" will be displayed.
2. When the value of the calibration weight begins to flash, load the calibration weight on the
plate.
-0-
3. Wait for the calibrated weight to be displayed and the stability symbol to turn on, and then
remove the weight from the plate.
4. When "0.000" is shown on the display, press the PRINT button in a prolonged manner until
the beeping stops. The acquisition and automatic storage of the internal weight will now
begin. During the acquisition cycle, the display will show the following flashing writing:
-0-
01-06-10
200
01-06-10
Mem. Tech
01-06-10
Cal
13:15:24
g
13:15:24
28
13:15:24