Installation Instructions - Accuris W3101A-120 Operation Manuals

Analytical balance series dx (graphic dot matrix display)
Table of Contents

Advertisement

1

Installation instructions

WARNING:
Please carefully read these installation and use instructions before you begin using
your new balance. The product's safety and accuracy cannot be guaranteed if the
instrument is used in a way that is different from that recommended in this manual.
DO NOT DISCARD THE ORIGINAL PACKAGING MATERIAL. THIS PACKAGING MUST BE
USED IF THE BALANCE NEEDS TO BE RETURNED FOR SERVICE.
▪ Remove the balance and its calibration weight (if supplied) from the carton and check for any
visible damage to the instrument.
▪ Do not install the balance in environments in which there are drafts, strong temperature
changes, or vibrations.
▪ Do not use the balance in potentially explosive environments.
▪ The humidity of the environment where the balance is used must be between 45% and
75%.
For models with internal calibration: remove the protection screw in the
left lower part of the balance (see Section 5)
ATTENTION
▪ Place the weigh pan support and weigh pan on the balance (see page 6).
▪ Level the balance by adjusting the front feet. The leveling bubble will indicate that the
balance is leveled properly.