A 25-seconds countdown will start during which you can decide if:
Stop the automatic-calibration procedure by pressing the "ON/OFF button that will be
delayed of 5 minutes
or
Let the countdown finish so that the automatic calibration starts
Note: during the countdown do NOT load anything on the weigh pan!
In this mode, it is also possible to carry out the calibration with the internal reference mass by
pressing the CAL button at any moment, first ensuring that no weight is loaded on the plate.
1. Press the CAL button with the plate empty. The display will show the message "CAL" and
the balance's calibration will be carried out automatically.
2. At the end of the calibration, the balance will return to normal weighing conditions.
If the calibration is not completed due to vibrations or drafts, the message "CAL bUt" will be
displayed. Press the CAL bUtton again, and if the problem persists, select external
calibration and contact the supplier.

9.12.2 Internal calibration (I-CAL) [locked for verified balances]

The balance calibrates itself through the internal reference mass ONLY upon the request of the
-0-
-0-
01-06-10
0.0000
-0-
01-06-10
13:15:24
Cal 25
01-06-10
Cal
13:15:24
g
27
13:15:24