2. At the end of the calibration, the balance will return to normal weighing conditions.
If the calibration is not completed due to vibrations or drafts, the message "CAL bUT" will be
displayed. Press the CAL button again, and if the problem persists, select external calibration
and contact the supplier.
To switch the calibration mode in these models to external calibration, see section 9.12.1

7.3 Verification

General introduction:
According to EU directive 90/384/EEC balances must be verified if they are used for legally
controlled applications:
a) For commercial transactions if the price of goods is determined by weighing.
b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory.
c) For official purposes.
d) For manufacturing final packages.
For any questions about requirements, please contact your local standards office.
Verification instructions
An EU type approval exists for balances described in their technical data as
verifiable. If a balance is used where obligation to verify exists as described above, it
must verified and re-verified in regular intervals.
Re-verification of a balance is carried out according to the respective national
regulations. The legal regulation of the country where the balance is used must be observed!
After verification the balance is sealed at the indicated positions.