10.5.2 Mode with manual insertion of the reference weight.

1. Select the percentage weight function as described in paragraph 10.
The following screen will be shown on the display:
Percent. weight
2. Press the MENU button to select manual mode
Percent. weight
3. Confirm manual mode by pressing the PRINT button.
4. You can now enter the reference weight value, using the CAL and MENU keys to increase
and decrease the value, while pressing the TARE button to pass to the next value. During
the entry phase, holding down the TARE button allows you to delete the value entered. The
value entered will remain in the memory until the balance is turned off. It is also possible to
enter the value using the optional alphanumeric keypad.
5. After having inserted the desired reference weight value, press the ENTER key.
6. Now load the sample and read the percentage value.
7. Press the ON/OFF button to exit from the percentage weighing function.
NOTE: If the reference weight entered is less than 10 displayed digits, the word ERROR 07 will be shown.
01-06-10
Automatic
Select the mode
01-06-10
Manual
Select the mode
Percent. weight 01-06-10
Enter the reference weight
Percent. weight 01-06-10
P: 0.0000 g
P%.
E premere ' Print'
13:15:24
13:15:24
13:15:24
0.0000
g
13:15:24
RIF: 50.0000 g
0.00 %
48