ABB CoriolisMaster FCM2000 Operating Instructions Manual page 85

Mass flowmeter
Hide thumbs Also See for CoriolisMaster FCM2000:
Table of Contents

Advertisement

Parameter
Function test
HART Transmitter
Function test
HART command
D184B111U02
Value range / input type
Qm
Measure
Enter
Enter
Enter
Qm
-115 ... +115 %
15.00 %
Technician*
Density
Measure
Enter
Enter
Enter
Technician*
Density
0.3 ... 3.7 g/ml
1.00 g/ml
Technician*
Temperature
Measure
Enter
Enter
Enter
Technician*
Temp. housing
Measure
Enter
Enter
Enter
Technician*
Temperature
-60 ... 190 °C
30 °C
Technician*
Temp. housing
Measure
20 °C
Enter
Enter
Technician*
Technician*
Technician*
CoriolisMaster FCM2000
Note
Table
This menu can be used to measure the actual
mass flowrate while in the simulation mode.
float
The mass flowrate value for the simulation can be
entered in percent. This menu is only displayed
when mass flowrate was selected. The allowable
range of entries is -115 % ... +115 %.
Table
This menu can be used to specify how to measure
the actual mass flowrate while in simulation mode.
float
The density value for the simulation can be
entered. This menu is only displayed when
density was selected. The allowable range of
entries is 0.3 ... 3.7 g/cm
Table
This menu can be used to specify how to measure
the temperature while in simulation mode.
Table
This menu can be used to specify how to measure
the temperature while in simulation mode.
float
The temperature value for the simulation can be
entered. This menu is only displayed when
temperature was selected. The allowable range of
entries is -60 ... 190 °C.
float
The two HART-Frequencies (1200 Hz and 2200
Hz) can be selected and transmitted.
The received HART commands are displayed.
Parameterization