ABB CoriolisMaster FCM2000 Operating Instructions Manual page 83

Mass flowmeter
Hide thumbs Also See for CoriolisMaster FCM2000:
Table of Contents

Advertisement

Parameter
TOT2 Channel
TB DiagExtMask
0x080001FE0FFF
CommSoftwareRev
2.11.0.12
Dip Switch
Submenu
Function test
Function test
Pulse Output
Function test
Iout 1
D184B111U02
Value range / input type
TB Mass Flow
TB Volume Flow
Bit-wise selection
of alarms
0 ... 126
Technician*
123456789A Addr.
- -x - x- -x x x 20
0.001 ... 5000 Hz
Technician*
Pulse Output
1 Hz
0 ... 26 mA
Technician*
Iout 1
10 mA
CoriolisMaster FCM2000
Note
Only contained in the PROFIBUS PA software.
This can be used to select the bits of the error
register to be transmitted. For details, please see
our FCM2000 data link documentation
PROFIBUS PA.
Only contained in the PROFIBUS PA software.
Indicates the communication software version.
Only contained in the PROFIBUS PA software.
Displays the switch position of the hardware
address switch for PROFIBUS PA after Enter is
pressed.
In this menu you can test inputs and outputs as
well as the individual components of the
converter. Also included is a Simulation menu,
which provides the ability to set all measurement
variables in the converter to programmable
values.
float
During normal operation the output of the pulses
is dependent on the flowrate selection (Mass or
Volume), the selected QmMax and the pulse
factor. In order to test the pulse output
independent of the converter settings, the self
check function permits the pulses per second
output on the pulse to be directly entered. The
value entered must be between 0.001 ... 5000
pulses/s. Pressing any key terminates the self
check process and the output is again a function
of the converter settings.
Not contained in the fieldbus software.
float
This function can be used to set a value for the
current output independent of the variable
assigned to the current output, for test purposes.
A current between 0 ... 26 mA can be simulated.
(In version A.00 the entry is made in percent)
Not contained in the fieldbus software.
Parameterization
83

Advertisement

Table of Contents
loading

Table of Contents