ABB CoriolisMaster FCM2000 Operating Instructions Manual page 80

Mass flowmeter
Hide thumbs Also See for CoriolisMaster FCM2000:
Table of Contents

Advertisement

Parameterization
Parameter
Contact output
No function
Submenu
Label
TAG number
1234ABCD
Descriptor
123456789ABCDE
Date
07.07.2001
Unit number
1234567
80
Value range / input type
No function
F/R-Signal _
F/R-Signal /
Contact output
General alarm _
Max Alarm_ _
General alarm /
MAX/MIN Alarm _
MAX/MIN Alarm /
MIN Alarm _
MIN Alarm /
MAX Alarm _
MAX Alarm /
8 ASCII
16 ASCII
1.1.1900 ...
31.12.2155
0 ... 999999
CoriolisMaster FCM2000
Note
Can be used to define the function of the contact
output. The following functions are available:
No function (contact opened)
F/R-Signal _ _
(if no forward flow → contact closed)
F/R-Signal /_
(if no forward flow → contact opened)
General alarm _ _
(if no general alarm → contact closed)
General alarm / _
(if no general alarm → contact opened)
MAX/MIN Alarm _ _
(if no MAX/MIN Alarm → contact closed)
MAX/MIN Alarm _ /
(if no MAX/MIN Alarm → contact opened)
MIN Alarm _ _
(if no MIN Alarm → contact closed)
MIN Alarm / _
(if no MAX/MIN Alarm → contact opened)
MAX Alarm _ _
(if no MAX/Alarm → contact closed)
MAX Alarm / _
(if no MAX Alarm → contact opened)
The symbols "/ _ " and "_ _" indicate normally
closed and normally open contacts. A normally
open contact closes when actuated (in this case
when the appropriate conditions are satisfied). A
normally closed contact operates in the reverse
manner.
ASCII
Not contained in the fieldbus software versions.
The TAG number can be used to identify the
meter location (HART Protocol). The length is
limited to 8 numbers according to HART
specification and/or capital letters (Packed ASCII).
HART descriptor
The length is limited to 16 numbers and/or capital
letters (Packed ASCII).
HART file
Display
D184B111U02