ABB CoriolisMaster FCM2000 Operating Instructions Manual page 37

Mass flowmeter
Hide thumbs Also See for CoriolisMaster FCM2000:
Table of Contents

Advertisement

Observe the following items when routing cables:
The signal cable carries a voltage signal of only a few millivolts and therefore must be routed
the shortest distance possible. The maximum permissible signal cable length is 50 m
(164 ft.).
Avoid routing the cable in the vicinity of electrical equipment or switching elements that can
create stray fields, switching pulses and induction. If this is not possible, run the signal cable
through a metal pipe and connect this to the station ground.
All leads must be shielded and connected to station ground.
Do not run the signal cable over junction boxes or terminal blocks.
To shield against magnetic interspersion, the cable contains outer shielding that is attached
to the SE clamp.
Fig. 33
1 Make sure during installation that the cable is provided with a water trap (1). For vertical
installation, align the cable glands pointing downward.
Pos: 31.2 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @
D184B111U02
1
CoriolisMaster FCM2000
Installation
G00413
37