Parameter
Temp → l = 100 %
180.0 °C
Qm% → l = 100 %
120 kg/min
Iout Alarm
Low
Low Alarm
3.2 mA
High Alarm
21.0 mA
Submenu
Switch contacts
Contact input
Totalizer reset.
D184B111U02
Value range / input type
-50 ... 180
float
Low
Table
High
Technician*
3.5 ... 3.6 mA
float
Technician*
21 ... 26 mA
float
Technician*
No function
Concentr. Table
Ext. output
Shut-off
Totalizer reset.
CoriolisMaster FCM2000
Parameterization
Note
In this menu the temperature value is entered for
which the current output is to indicate its 100 %
value (20 mA). The menu is only displayed when
the current output is based on the temperature.
Minimum:
-50 °C
Maximum:
180 °C
The temperature value for a current output value
of 100 % must be at least 10 °C larger than the
temperature value for a current output value of
0 %.
Only displayed for "Output of Qm Conc." Net
massflow at current output 100 %
In this menu a selection can be made if the
current output during an alarm condition is to be
set to High-Alarm-Current or the Low-Alarm-
Current. For some error conditions the current
output, independent of the selection, is always set
to the High-Alarm-Current or to the Low-Alarm-
Current (see Alarm Overview).
In this menu the value of the Low-Alarm-Current
can be entered. The Low-Alarm-Current can be
set between the limits of 2 ... 3.6 mA.
In this menu the value of the High-Alarm-Current
can be entered. The High-Alarm-Current can be
set between the limits of 21 mA and 26 mA.
In this submenu the function assigned to the
contact input and the contact output can be
defined.
Can be used to define the function of the contact
input. The following functions are available:
No function
Concentration Table
with selection "Conc. Table" variable matrix 1
or matrix 2 can be selected.
Ext. zero return
(Current and pulse outputs are set to 0%
flowrate. Internal totalizers are halted.)
Totalizer reset.
(resets all mass and volume flow totalizers)
Not contained in the fieldbus software versions.