ABB CoriolisMaster FCM2000 Operating Instructions Manual page 84

Mass flowmeter
Hide thumbs Also See for CoriolisMaster FCM2000:
Table of Contents

Advertisement

Parameterization
Parameter
Function test
Iout 2
Function test
Contact input
Function test
Contact output
Function test
Memory
Submenu
Simulation
84
Value range / input type
on
off
Technician*
Iout 2
10 mA
on
off
Technician*
Contact input
on
on
off
Technician*
Contact output
on
Technician*
Technician*
Simulation
on
on
off
Technician*
CoriolisMaster FCM2000
Note
float
This function can be used to set a value for the
current output independent of the variable
assigned to the current output, for test purposes.
A current between 3.5 ... 26 mA can be simulated
Not contained in the fieldbus software.
Table
This function can be used to determine whether
the contact input is opened of closed.
Not contained in the fieldbus software.
Table
Can be used to manually toggle the contact
output.
Not contained in the fieldbus software.
The submenu Simulation contains a number of
menus with which it is possible to set individual or
even all measurement variables of the converter
to programmable values.
Table
When the Simulation is turned on additional
menus are available in the submenu Simulation.
They enable you to determine for every variable
whether the variable should be measured or
simulated and what value should be assigned to
the variable. Therefore a selection can be made
from the following:
Measure
→ The measurement variable is measured
Enter
→ The measurement variable can be simulated.
A fixed value can be entered in the
appropriate menu
Enter
→ The simulated value entered can be
increased or decreased in steps using the
STEP- and DATA- keys if the process display
is shown.
Simulated values for the variables can be entered
which exceed the alarm limits to test the operation
of the alarms.
The settings in the simulation menu are not
stored. After a power interruption all simulation
menus are turned off.