2 Installing the Bravo Platform
Starting up and shutting down
Shutting down the Bravo Platform
Related information
50
Shut down the Bravo Platform before you:
Clean the Bravo Platform
Change the liquid- handling head
Install accessories
Move the Bravo Platform
CAUTION
Always turn off the Bravo Platform before installing or uninstalling a liquid-
handling head. Failure to do so can damage the head electronics.
To shut down the Bravo Platform:
1
If removing the liquid- handling head, ensure that the disposable tips are
removed.
2
Optionally, home the liquid- handling head.
3
Shut down the computer.
4
Turn off any accessories, for example, Pump Module.
5
If using an Auto Filling Reservoir, disconnect the bottles to prevent
siphoning.
6
On the side of the Bravo Platform, press the power switch to the off (o)
position.
7
If applicable, uninstall the liquid- handling head.
For information about...
Setting up and using the Bravo
Platform
See...
Bravo Platform User Guide
For details on how to access this
guide, see
information" on page
G5562A, G5563A Bravo Platform Safety and Installation Guide
"Where to find user
vii.