G5562A, G5563A Bravo Platform Safety and Installation Guide
4
Remove the head from the stand as follows:
Disposable- tip heads. Rest the bottom of the stand on a clean, dry
surface. Slide the head out of the stand as the following figure shows,
so that the barrels are facing down.
Figure Disposable-tip head: (1) retainer pins and (2) stand
Bravo 96AM Head. Ensure that the top of the head is resting on a
clean, stable surface so that the probes are facing up.
WARNING
The probes of the AssayMAP Bravo 96AM Head are sharp and can
scratch you if they brush across your hand. A probe scratch can expose you to any
contaminants remaining on the probes. Wear gloves and use extreme caution to
avoid brushing against the probes.
Using both hands, carefully lift the stand off of the head while guiding
the stand's side cutouts (1) off the head side tabs (2). Use care to avoid
touching the probes.
1
2
2 Installing the Bravo Platform
Installing the liquid-handling head
25