Agilent G4560A Upgrade Kit For the Agilent 7697A Headspace Sampler Installation Guide Parts Supplied The Agilent 7697A Upgrade Kit comes in three boxes described in Table Table 2, and Table Table 1 Box 1 (labeled G4556-67009) Description Quantity 111 Vial Model Oven Assembly...
Page 6
Agilent G4560A Upgrade Kit Remove covers from the headspace sampler Remove the covers shown in Figure 4 from the headspace sampler. Pneumatics front cover Top cover Top cover tray Rotating top cover Right panel assembly Front cover Figure 4 Covers to remove from the 12 vial model headspace sampler...
Page 7
Agilent G4560A Upgrade Kit Remove the rotating top cover (Figure Rotating top cover Thumbscrews Figure 5 Removing the rotating top cover Lift the rotating top cover from the front so that the two thumbscrews are accessible underneath. Unhinge and remove the tray cover by pulling one of the two thumbscrews inward while lifting the tray cover off of the chassis.
Page 8
Agilent G4560A Upgrade Kit Remove the pneumatics front cover. (See Figure Remove the T- 20 screw that secures the cover in place (Figure Figure 6 T-20 Torx screw securing the pneumatics front cover Slide out, then lift to remove the pneumatics front cover.
Page 9
Agilent G4560A Upgrade Kit Remove the pneumatics assembly from the headspace sampler. Completely loosen the two T- 20 Torx captive thumbscrews that are inset at the front of the pneumatics assembly (Figure Figure 7 Front T-20 Torx thumbscrews Completely loosen the T- 20 Torx captive thumbscrew in the back...
Page 10
Agilent G4560A Upgrade Kit the electronics control board for easy re- installation. Disconnect the 12- vial model heater assembly cables from the headspace sampler. Disconnect the oven heater cable (Figure 9). The oven heater connection jack is located on the left side behind the oven compartment facing the front of the headspace sampler.
Page 11
Agilent G4560A Upgrade Kit Loosen the four T- 20 Torx screws shown in Figure Figure 11 Loosening four T-20 Torx screws If any broken glass exists, carefully remove it from the carousel and vial tray area using a vacuum cleaner.
Page 12
Agilent G4560A Upgrade Kit Adjust the oven side racks on the headspace sampler The oven side racks are located just below the oven compartment on either side of the headspace chassis as shown in Figure Oven side racks Figure 12 Locating the oven side racks...
Page 13
Agilent G4560A Upgrade Kit Left side oven rack Figure 13 Removing the T-20 Torx screw on the left side oven rack Right side oven rack Figure 14 Removing the T-20 Torx screw on the right side oven rack Installation Guide...
Page 14
Agilent G4560A Upgrade Kit Carefully push the left side rack towards the back of the headspace sampler (Figure 15). Figure 15 Sliding the side rack (left side oven rack shown) Remove the left side rack by pulling it out and off of the support rivets.
Page 15
Agilent G4560A Upgrade Kit Install the 111 vial model oven assembly The oven assembly fits tightly into the headspace oven area. To avoid injury, wear WA R N I N G gloves and be cautious of sharp edges when placing the oven assembly into the headspace sampler.
Page 16
Agilent G4560A Upgrade Kit Connect the ribbon cable to connection P1 on the control board (Figure 18). Figure 18 Connect the ribbon cable to P1 on the control board Route the oven sensor cable through the insulated opening in the...
Page 17
Agilent G4560A Upgrade Kit Route the cables from the oven assembly through the opening in the headspace chassis and to the main board as shown in Figure Figure 20 Routing the oven cables to the main board Be sure that the cables run cleanly from the oven assembly to the main board.
Page 18
Agilent G4560A Upgrade Kit Connect the oven heater cable to P22 on the main board (Figure 21). Figure 21 Connecting the oven sensor cable to P22 on the main board Install the 111 vial model top instrument cover Install the pneumatics assembly.
Page 19
Agilent G4560A Upgrade Kit Remove the T- 20 Torx screw from the headspace chassis bracket found to the right of the oven assembly and behind the display panel (Figure 22). This screw will be reused in step Figure 22 Remove the T-20 Torx screw from the headspace chassis bracket Place the top instrument cover onto the headspace.
Page 20
Agilent G4560A Upgrade Kit • Be sure that the rivet shown in Figure 24 passes entirely through the opening on the top instrument cover. Figure 24 Top instrument cover correctly installed Route and connect the tray cables to connections P41, P42, and P43 on...
Page 21
Agilent G4560A Upgrade Kit Install the T- 20 Torx screws shown in Figure 26 to secure the top instrument cover to the headspace. Figure 26 Install T-20 Torx screws to secure the top instrument cover to the headspace Installation Guide...
Page 22
Agilent G4560A Upgrade Kit Install 111 vial model tray Lower the back of the tray assembly onto the top instrument cover while holding the front of the tray assembly above the headspace. Slide the cables into the tray cable clip (Figure 27).
Page 23
Agilent G4560A Upgrade Kit Connect the tray cables to the cable connections (Figure 28). Cable in clip Cable connections Figure 28 Tray cable connections Lower the front of the tray assembly onto the headspace. Be sure it is sitting level on the headspace sampler.
Page 24
Agilent G4560A Upgrade Kit Install the 111 vial model covers Install the right panel assembly. Lower the bottom of the panel into the headspace sampler and slide the panel assembly into position. Install the two T- 20 Torx screws on the right panel assembly.
Page 25
Agilent G4560A Upgrade Kit Power on the headspace sampler Power on the headspace sampler (Figure 29). Figure 29 Power on the headspace sampler Press [Status] on the headspace keypad. The headspace status displays a message similar to what is shown in...
Page 26
Agilent G4560A Upgrade Kit Select the Agilent 7697A Headspace Sampler from the Instruments list and click Connect. If the Agilent 7697A Headspace Sampler is not available, click Add New Instrument to open Configure Instruments. Add the new instrument, then return to firmware update.
Page 27
Agilent G4560A Upgrade Kit Select the Show detailed version information checkbox (Figure 31). Figure 31 Showing detailed version information in the Firmware Update utility In the HCV App row, select the same firmware revision as is installed for the LCV App (Figure 32).