2 Installing the Bravo Platform
Laboratory setup requirements
Electrical requirements
Environmental operating requirements
Computer requirements
16
The Bravo Platform has the following electrical requirements:
Requirement
Voltage
Frequency
Power, maximum
Fuse
The Bravo Platform is for indoor use only. The following table lists the
operating specifications.
If you have integrated devices, your system might require additional cooling
depending on the number and types of integrated devices.
Operating
Pollution degree
Installation category
Temperature
Humidity
Altitude
Ensure that the computer contains the following software versions at a
minimum.
Microsoft Windows 10 64- bit operating system
VWorks Automation Control software v13.1
See the VWorks software release notes for installation instructions and
additional computer requirements. You can also go to the VWorks knowledge
base at www.agilent.com/chem/askb.
Value
100–240 V~
50/60 Hz
300 VA
250 V, 6.3 A, 5 mm x 20 mm, time delay
Specification
2
II
4–40 °C
10–90% RH, non- condensing
1–2000 m
G5562A, G5563A Bravo Platform Safety and Installation Guide