2 Installing the Bravo Platform
Installing the Standard Light Curtain and shields
Installing the front shields
42
Figure Adjustment screws (1) on the lightpost interior top and bottom
6
To verify the adjustment:
a
Press the Bravo power switch to the off (o) position, wait a minute, and
then press the power switch to the on (|) position.
b
Verify that the entire column of LEDs on each lightpost are solid green
and not blinking.
Before you start
Remove the protective backing from the shields.
If necessary, use a soft, lint- free cloth and isopropyl alcohol to clean the
shields.
To install the front shields:
1
At the front of Bravo Platform, position the upper front shield so that the
side with the countersink screw holes is facing away from the device.
2
Secure the upper front shield to the top of the two lightposts using the
four screws provided.
3
Position the lower front shield so that the side with the countersink screw
holes is facing away from the device.
4
Secure the lower front shield to the two lightposts using the four screws
provided.
After installation, the flat screw heads should be flush with the shield surface.