2 Installing the Bravo Platform
Bravo light curtains and shields
Standard Light Curtain and shield components
Item
Name
1
Lightposts
2
Upper front
shield
3
Lower front
shield
4
Light shield
(invisible)
30
The Standard Light Curtain contains two lightposts mounted at the front of the
Bravo Platform that project light beams across the front of the device. If an
object disrupts the light beams, the safety interlock circuit is tripped and the
motion of the Bravo head stops.
The following figure and table provide details on the Standard Light Curtain
and the front and side shields.
For installation instructions, see
shields" on page
36.
Figure Bravo Platform with Standard Light Curtain and shields (front view)
4
Description
The two lightposts, which mount on each end of the Bravo base,
contain the transmitter and receiver that project the light beams, detect
disruptions in the beam, and transmit signals to the safety interlock
circuit.
A clear plastic shield, which attaches to the two lightposts, prevents
access at the front between the top of the light shield and the Bravo
top.
A clear plastic shield, which attaches to the two lightposts, prevents
access at the front between the bottom of the light shield and top of
the Bravo deck.
After the light beams are aligned, the Standard Light Curtain generates
an invisible light shield that spans the area between the two lightposts
across the front of the Bravo Platform.
"Installing the Standard Light Curtain and
G5562A, G5563A Bravo Platform Safety and Installation Guide