2 Installing the Bravo Platform
Connecting the Bravo Platform
Pump I/O port
20
Item
Feature
2
Pendant port
3
Fuse holder
4
AC power
entry
The following figure shows the Pump I/O port on the back of the Bravo
Platform. The table below describes the Pump I/O port in detail.
Figure Pump I/O port (Bravo Platform back view)
00212c
Bravo
Acc_IO
Item
Feature
1
Pump I/O port
Description
Connects the emergency- stop pendant and a Bravo
light curtain to the Bravo safety interlock circuit.
The connection is made through the junction box
for a standalone Bravo Platform. For an integrated
Bravo Platform, the connection is made through a
main emergency- stop control, such as the Agilent
Automation Control Unit or Robot Disable Hub.
IMPORTANT
The safety interlock circuit must be
closed for the Bravo Platform to operate.
Houses the AC inlet fuse and a spare fuse for the
Bravo Platform.
Connects the Bravo power cord to an AC outlet
with a grounded circuit.
1
Description
An RJ- 45 I/O port that connects the Pump
Module to the Bravo Platform. The connection is
made with a straight- through shielded Cat- 5 or
Cat- 6 (Ethernet) cable.
IMPORTANT
This is not an Ethernet port.
G5562A, G5563A Bravo Platform Safety and Installation Guide