Agilent Technologies G5562A Safety And Installation Manual page 46

Bravo series automated liquid-handling platform
Hide thumbs Also See for G5562A:
Table of Contents

Advertisement

2 Installing the Bravo Platform
Installing the Standard Light Curtain and shields
Connecting the Standard Light Curtain
38
Figure Lightpost setscrew (Bravo Platform side view)
3
Repeat steps 1 and 2 to mount the front right lightpost.
Use the following procedure to connect the Standard Light Curtain to a
standalone Bravo Platform using the junction box.
If the Bravo Platform is integrated with an Automation Control Unit, see the
Automation Control Unit User Guide
If the Bravo Platform is integrated with a Robot Disable Hub, see the
Microplate Handler User Guide
The ports for the pendant cable and the two lightpost cables are on the front
of the junction box. The Bravo port for the extension cable is on the back of
the junction box. The following figure shows the junction box connections.
Figure Junction box connections (front view and back view)
PENDANT
RX
To use a junction box to connect the Standard Light Curtain:
Refer to the following figure and table to connect the junction box to the
Standard Light Curtain.
1
for connection instructions.
for connection instructions.
TX
G5562A, G5563A Bravo Platform Safety and Installation Guide
BenchCel
B R AVO

Advertisement

Table of Contents
loading

This manual is also suitable for:

G5563a

Table of Contents