Printout for Differential Weighing
Generating Printouts
Automatically
The configured backweighing printout
is generated automatically after
backweighing, if one of the following
settings is selected Setup: Application 1:
Differential weighing: Generate print-
out:
Automatic after
backweighing
Auto after init.weigh
+ backweigh
Auto after tare-,
init- +backweigh.
Generating Printouts Manually
The individual printout is generated
when the r key is pressed while
there is a tare, initial or backweight on
the balance, or when d is pressed
to toggle applications.
You can generate the configured
printout manually after backweighing
if you press the r key while the
display page for the results is shown.
To generate the statistics printout,
press the r key
– when the display page for statistics
is shown,
– when the samples with a desired number
of backweighing operations is selected
(for example, statistics on all samples
with 2 backweighing operations).
The following printout is generated:
Backweighing Printout
(Example)
--------------------
16.11.1999
14:55:12
Lot
CH12345
Sample
ID
CX88
T1
+ 23.45821 g
N1
+125.57234 g
R
(3)+103.68442 g
R
+
82.57 %
D
- 21.88792 g
D
-
17.43 %
Fact
+
1.10345
D-Res -
24.15 o
Ratio1+
21.11 %
Ratio2+
121.11 %
--------------------
Dotted line
Date/time

Lot ID

14
Sample number
Sample ID
Tare weighing (with PT1 selected)
Initial weight
Backweight (residue as weight)
Residue in percent
Loss as a weight
Loss in percent
Calculation factor
Calculated loss
Ratio 1
Ratio 2
Dotted line