Sartorius ME215 Operating Instructions Manual page 100

Me series se series electronic analytical balances and semi-micro-, micro- and ultra-microbalances
Table of Contents

Advertisement

Example
Determination of the Air Density by Weighing, Using the Reference Weights Supplied for Steel and Aluminum
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 1: Air buoyancy correction: Air buoyancy determination: Yes
Step
1. Turn on the balance if it is
not already on
2. Start air buoyancy correction
3. Toggle to the mode for air density
determination to enter the
specifications for steel and aluminum
4. Toggle to the display of the
air density parameters
5. Toggle to the display of the
air density references
6. Enter the specification of the steel
reference supplied (in this case,
3
200.00821 g/cm
)
7. Confirm entry
8. Enter the specification of the
aluminum reference supplied
(in this case, 200.00348 g/cm
100
Air buoyancy correction: Change steel/aluminum references: Yes
Press key(s) (or follow instructions)
e
Start
soft key
RhoA
soft key
Param.
soft key
Ref.
soft key
200.
00821
soft key
l
200.
00348
3
)
Display/Output

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Me235Me254Me36sMe5Se2Me414 ... Show all

Table of Contents