Sartorius ME215 Operating Instructions Manual page 141

Me series se series electronic analytical balances and semi-micro-, micro- and ultra-microbalances
Table of Contents

Advertisement

Output Format (Line Format)
You can output the values displayed in the
line for measured values and the weight
unit with or without a data ID code
Example: Without data ID code
+
253 pcs
Example: With data ID code
Qnt
+
253 pcs
Configure this parameter in the Setup
menu (Setup: Printout: Line format).
The output with a data ID code has
16 characters; without a data ID code,
22 characters.
Output Format With 16 Characters
Display segments that are not activated
are output as spaces. Characters without
a decimal point are output without a
decimal point.
The following characters can be output, depending on the characters displayed on the balance:
Normal Operation
Position
1
2
3
4
5
+
D
D
D
D
or
.
.
.
.
or
*
*
*
*
*
*:
Space
D:
Digit or letter
U:
Unit symbol
CR:
Carriage return
LF:
Line feed
Special Codes
Position
1
2
3
4
5
*
*
*
*
*
or
*
*
*
*
*
or
*
*
*
*
*
and only upon request with ESC w0 (no print command):
*
*
*
*
*
or
*
*
*
*
*
*:
Space
AB = – –:
Final readout mode
A = H:
Overload
AB = H H:
Overload in checkweighing
A = L:
Underload
AB = L L:
Underload in checkweighing
XXX =
Decimal value calculated from individual status bits:
Decimal
Binary
value
value
Bit0 = 0:
1
Bit0 = 1:
Bit1 = 0:
2
Bit1 = 1:
Bit3 = 0:
8
Bit3 = 1:
Bit4 = 0:
16
Bit4 = 1:
Bit6 = 0:
64
Bit6 = 1:
Example for ME215/235/254/414/614/36S:
R, M, L = COO: right door closed, middle and left doors open
R, M, L = OCC: right door open, middle and left doors closed
Example for ME5, SE2:
W 008210
Angle of aperature:
The draft has turned itself by 210°.
Control information:
0
Bit 6 – Motorized draft shield operation
+ 0
Bit 4 – Draft shield open
+ 8
Bit 3 – Learning function on
+ 0
Bit 1 – Draft shield motor off
+ 0
Bit 0 – No error
= 08
6
7
8
9
10 11 12 13 14 15 16
D
D
D
D
D
*
U
.
.
.
.
.
*
*
*
*
*
*
6
7
8
9
10 11 12 13 14 15 16
*
*
*
*
*
*
A
*
*
*
*
*
*
A
B
*
*
*
*
*
S
*
X
X
X
Y
*
I
*
X
X
X
*
A = C:
Cal/adjustment
S:
Draft shield status
I:
Ionizer
Y,Y,Y =
Draft shield doors
Control
information
No error/ionizer off
Draft shield error/ionizer on
Draft shield motor off
Draft shield doors in motion
Learning function off
Learning function on
At least one draft shield door open
All draft shield doors closed
Motorized draft shield operation
Manual draft shield operation
U
U CR LF
*
*
*
*
CR LF
*
*
CR LF
*
*
CR LF
Y
Y
CR LF
*
*
CR LF
141

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Me235Me254Me36sMe5Se2Me414 ... Show all

Table of Contents