Care and Maintenance
The Avant 9600 requires no calibration or periodic maintenance other than battery replacement.
The device's expected service life is 5 years.
Do not attempt to open the case or repair the electronics. Opening the case will damage the
device and void the warranty. If the device is not functioning properly, see "Troubleshooting."
The Oxitest
Plus7
by Datrend Systems, Inc. can be used to verify operation of the pulse oximeter.
Cleaning the Model 9600
Clean the Avant 9600 with a soft cloth dampened with isopropyl alcohol. Do not pour or spray
any liquids onto the device, and do not allow any liquid to enter any openings in the device. Allow
the device to dry thoroughly before reusing it.
CAUTION: Do not autoclave or immerse this device in liquid or use caustic or
!
abrasive cleaning agents.
Refer to the pulse oximeter sensor Instructions for Use for cleaning information.
26
Care and Maintenance