Nurse Call

Voltage
Current
Output Impedance
Ground Reference
Electrical Isolation
Output
Output Connector Type
Serial Port Pin-out
PIN #
1
2
3
4
5
6
7
8
9
40
30 VAC or DC (non-polarized), maximum
100mA continuous (maximum)
30 ohms (maximum)
Isolated ground
1500 VDC
Normally Open or Normally Closed; switch selectable
0.141" phone jack
RS-232 FUNCTION
Device Carrier Detect
Received Data
Transmitted Data
Data Terminal Ready
Signal Ground
Data Set Ready
Request to Send
Clear to Send
Ring Indicator