Nurse Call Feature; Table 6. Option Switches - Nonin Avant 9600 Operator's Manual

Hide thumbs Also See for Avant 9600:
Table of Contents

Advertisement

Switch
Switch 1
Alarm Disable
Up—Alarm volume may be disabled
Down—Alarm volume cannot be set to zero (disabled)
Switch 2
Normal / Slow SpO
Up—Slow Averaging (8 beat exponential average)
Down—Normal Averaging (4 beat exponential average)
Switch 3
Alarm Unlocked / Locked
Up—Alarms Locked
Down—Alarms Unlocked
Switch 4
Factory / User-Defined Defaults
Up—User-Defined Defaults for Alarm Limits and Volume Settings
Down—Factory Defaults for Alarm Limits and Volume Settings
Switch 5
US / International Date Format
Up—International Date Format
Down—US Date Format
Switch 6
Power Saving During Battery Use (Dims Display)
Up—Power Saving Disabled
Down—Power Saving Enabled
Switch 7
Nurse Call Output
Up—Continuous
Down—One second audible alarm at 60 second intervals
Switch 8
Serial Data Output
Up—Fast SpO
Down—SpO
switch 2

Nurse Call Feature

The Avant 9600 features a nurse call circuit that can be connected to a facility nurse call system,
allowing alarm conditions to be recognized at a central monitoring location and on the
Avant 9600. Two switches located on the back of the 9600 under the back cover allow users to
activate the Nurse Call feature.
14
Table 6: Option Switches
Averaging
2
and pulse rate output
2
and pulse rate output, as determined by the position of option
2
Function
Operating the Avant 9600