Front Panel Buttons - Nonin Avant 9600 Operator's Manual

Hide thumbs Also See for Avant 9600:
Table of Contents

Advertisement

Displays, Indicators, and Controls
Alarm Volume LED
This amber LED indicates that the Alarm Volume can be reviewed or
adjusted using the Plus (+) and Minus (-) buttons.
AC Power Supply LED
This green LED is displayed when an external power supply is providing
power to the Avant 9600. Any time this LED is displayed, the battery pack
is charging.
NOTE: When the external power supply is disconnected, the device
automatically switches to battery power without loss of functionality.
Battery LED
This amber LED indicates marginal battery capacity when blinking. When
solidly lit, the LED indicates that the battery pack is not installed. This LED
does not indicate that the device is running on battery power. New battery
packs must complete one charge/discharge/recharge cycle before the
Battery LED displays the actual capacity.
WARNING: The battery pack must be installed at all times while the device is
operating—even when operating on AC power. The audible alarms and memory will
not function if batteries are removed from the device. DO NOT use this device without
batteries.

Front Panel Buttons

ON/STANDBY
• Press Once: Turns the device on
• Hold 1 second: Turns the device off
• Hold 5 seconds: Turns the device off when in Patient Security mode
• Press momentarily while the unit is on:
• Indicates the battery capacity in green, in 12% increments for
• Initiates Print-on-Demand feature
• Records event marker
Time/Date
This button displays the time and date. Year, month, day, hour, and minute
can be set using the Plus (+) and Minus (-) buttons.
3 seconds
7