ferrotec CARRERA Operation Manual page 21

High-voltage power supply unit
Table of Contents

Advertisement

High Voltage Power Supply
Type: CARRERA
Version 1.3_15 EN
The switching and safety facilities for setting up, testing,
shutting down (including shutting down in an emergency),
operation, maintenance, cleaning and servicing are to be
installed and verified by the person setting up the entire unit.
This includes also lockable main switches and emergency
shut off switches
The owner must ensure that the operating stations are not
affected by waste gases (such as diesel engine exhaust
fumes) or a lack of oxygen.
The workplaces are to be assessed by the owner in
accordance with the noise and vibration regulations. Any
noisy areas are to be marked, the employees are to be
instructed about them and must use hearing protection.
The owner is required to provide suitable safety measures for
protection against fire such as suitable fire extinguishers of
the stipulated number and size at easily accessible points
and is to instruction his employees about protection against
fire
The warning notes from the documentation for modules
supplied by other companies are to be complied with and
integrated into the workplace-related hazard evaluations
The operator is required to ensure that the high voltage
power supply is only operated when it is in perfect working
order
Connections:
Before starting the high voltage power supply the operator must
ensure that:
only specifically trained personnel are used for installation
-
and start-up if this work is done by the operator himself and
that local rules and laws are followed in the way this is
installed..
Lighting
The operator must ensure that there is adequate and uniform
lighting in all areas of use of the electron beam evaporator. The
recommended illumination is 300 Lux (maintenance value;
ASR 7/3 applies in Germany).
NOTICE
20