Flow Connections - Agilent Technologies G1329A User Manual

1260 infinity series. preparative autosampler
Hide thumbs Also See for G1329A:
Table of Contents

Advertisement

3
Installing the Autosampler

Flow Connections

Flow Connections

Parts required
#
1
Preparations
• Sampler is installed in the LC system
Toxic and hazardous solvents
WA R N I N G
The handling of solvents and reagents can hold health risks.
1 Connect the pump outlet capillary to port 1 of the injection valve.
2 Connect column-compartment inlet capillary to port 6 of the injection
3 Connect the corrugated waste tube to the solvent waste from the leak plane.
4 Ensure that the waste tube is positioned inside the leak channel.
Do not extend the waste capillary of the autosampler. The siphoning effect might empty the
N O T E
complete seat capillary introducing air into the system.
48
Description
Parts from the Accessory kit
When opening capillary or tube fittings solvents may leak out.
Please observe appropriate safety procedures (for example, goggles, safety gloves
and protective clothing) as described in the material handling and safety data sheet
supplied by the solvent vendor, especially when toxic or hazardous solvents are
used.
valve.