Flow Connections
Preparation
Parts required
When opening capillary or tube fittings, solvents may leak out. Please observe
WA R N I N G
appropriate safety procedures (for example, goggles, safety gloves and protective
clothing) as described in the material handling and safety data sheet supplied by the
solvent vendor, especially when toxic or hazardous solvents are used.
1 Connect the pump outlet capillary to port 1 of the injection valve.
2 Connect column-compartment inlet capillary to port 6 of the injection
3 Connect the corrugated waste tube to the solvent waste from the leak plane.
4 Ensure that the waste tube is positioned inside the leak channel.
Do not extend the waste capillary of the autosampler. The siphoning effect might empty the
N O T E
complete seat capillary introducing air into the system.
Agilent 1200 Series Autosampler Reference Manual
Sampler is installed in the LC system
Parts from the Accessory kit
valve.
Installing the Autosampler
1
33