Performance Specifications - Agilent Technologies G1329A User Manual

1260 infinity series. preparative autosampler
Hide thumbs Also See for G1329A:
Table of Contents

Advertisement

2
Site Requirements and Specifications

Performance Specifications

Performance Specifications

Table 5
Type
Pressure
GLP features
Communications
Safety features
Injection range
Replicate injections
Precision
Minimum sample volume
Carryover
Sample viscosity range
Replicate injections per vial 1 – 99
Sample capacity
Injection cycle time
28
Performance Specifications Agilent 1200 Infinity Autosampler (G1329A). Valid
when standard 100 µl metering head installed.
Specification
Operating range 0 – 40 MPa (0 – 400 bar, 0 – 5900 psi)
Early maintenance feedback (EMF), electronic records of
maintenance and errors
Controller-area network (CAN). GPIB (IEEE-448), RS232C,
APG-remote standard, optional four external contact closures and
BCD vial number output
Leak detection and safe leak handling, low voltages in maintenance
areas, error detection and display
0.1 – 100 µl in 0.1 µl increments Up to 1500 µl with multiple draw
(hardware modification required)
1 – 99 from one vial
< 0.25% RSD from 5 – 100 µl, < 1% RSD 1 – 5 µl
variable volume
1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample in
300 µl microvial
Typically < 0.1%, < 0.05% with external needle cleaning
0.2 – 50 cp
100 × 2-ml vials in 1 tray
40 × 2-ml vials in ½ tray
15 × 6-ml vials in ½ tray (Agilent vials only)
Typically 50 s depending on draw speed and injection volume