Site Requirements and Specifications

Site Requirements

Site Requirements

A suitable environment is important to ensure optimum performance of the
instrument.

Power Consideration

The autosampler power supply has wide-ranging capability (see
page 27). Consequently there is no voltage selector in the rear of the
autosampler. There are also no externally accessible fuses, because automatic
electronic fuses are implemented in the power supply.
The thermostatted autosampler comprises two modules, the standard or
preparative autosampler and the thermostat (G1330B). Both modules have a
separate power supply and a power plug for the line connections. The two
modules are connected by a control cable and both are turned on by the
autosampler module.
Damaged electronics
WA R N I N G
Disconnecting or reconnecting the sampler to thermostat cable when the power
cords are connected to either of the two modules will damage the electronics of the
modules.
Incorrect line voltage at the instrument
WA R N I N G
Shock hazard or damage of your instrumentation can result, if the devices are
connected to a line voltage higher than specified.
24
Make sure the power cords are unplugged before disconnecting or reconnecting the
sampler to thermostat cable.
Connect your instrument to the specified line voltage.
Table 4
on

Hide quick links: