Installing the Manual Injector

Flow Connections

When opening capillary or tube fittings, solvents may leak out. Please observe
WA R N I N G
appropriate safety procedures (for example, goggles, safety gloves and protective
clothing) as described in the material handling and safety data sheet supplied by the
solvent vendor, especially when toxic or hazardous solvents are used (see also
"Leak Drainage"
1 Connect the pump outlet capillary to port 2.
2 Connect the column-compartment inlet capillary to port 3.
3 Connect the sample loop between ports 1 and 4.
4 Connect one vent capillary (supplied with valve) to port 5 and one to port 6.
The outlets of the two vent capillaries (ports 5 and 6) and the needle port must be at
C A U T I O N
the same level to prevent siphoning (see
14
on page 16).
Figure 6
on page 15).
1200 Series MI Reference Manual