Needle Assembly - Agilent Technologies G1329A User Manual

1260 infinity series. preparative autosampler
Hide thumbs Also See for G1329A:
Table of Contents

Advertisement

7
Maintenance
Simple Repairs

Needle Assembly

When
When the needle is visibly damaged
When the needle is blocked
Tools required
• ¼ inch wrench (supplied in accessory kit)
• 2.5 mm Hex key (supplied in accessory kit)
• A pair of pliers
Parts required
#
1
1
1
1
Preparations
• Select "Start" in the maintenance function "Change Needle".
• When the needle is positioned approx.15 mm above the needle seat, remove the front cover.
Personal injury
WA R N I N G
To avoid personal injury, keep fingers away from the needle area during autosampler
operation.
86
Part number
Description
G1313-87201
Needle assembly for G1313-87101 or G1313-87103 needle-seat
G1329-80001
Needle assembly for G1329-87101 or G1329-87103 needle seat
G1313-87202
Needle assembly (900 µl loop cap) for G1313-87101 needle seat
G2260-87201
Needle assembly (900 µl loop cap) for G2260-87101 needle-seat
Do not bend the safety flap away from its position, or attempt to insert or remove a
vial from the gripper when the gripper is positioned below the needle.